SPELLMAN高压电源广泛应用于东芝CT机XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下

Error contents (note 1)


Cause of problem

Object part




24kW

48kW/60kW

E13

Arc-

Cathode side Arc of the X-ray tube or the high-voltage section

TUBE, INV/HV

TUBE, INV/HVC

E25

Arc+

Anode side Arc of the X-ray tube or the high-voltage section


TUBE, INV/HVA

E12

Over kV-

Cathode side overvoltage. Abnormal main circuit control.

INV/HV, SCB

INV/HVC, SCB

E27

Over kV+

Anode side overvoltage. Abnormal main circuit control.


INV/HVA, SCB

E62

Over mA-

Cathode side overcurrent of the X-ray tube or the high-voltage section.

TUBE, INV/HV, SCB

TUBE, INV/HVC,SCB

E63

Over mA+

Anode side overcurrent of the X-ray tube or the high-voltage section.


TUBE, INV/HVA,SCB

E66

MAC error

Filament current abnormal. Trouble of the filament heatig circuit.

INV/HV, SCB

INV/HVC, SCB

E0

Regulation error (HV abnormal)

There is a difference in binding output and real output

INV/HV, AC/C

INV/HVC, INV/HVA, AC/C